Dwyer GFM2 Series Features
- Utilizes a straight tube sensor with a restrictor flow element to provide a high ±1% FS accuracy and ±0.25% FS repeatability
- Gas flow can be displayed in 23 different engineering units on an optional 2x16 character LCD display with internal conversion factors for up to 32 gases
- Digital RS-232 or RS-485 interfaces allow for easy communication and for multi-drop capability of up to 256 units (RS-485 only)
- User-friendly interface allows for the programming of high and low gas flow alarms, along with two user-programmable electromechanical SPDT relays with latch options
- Stores calibration information for up to 10 different gases, internal or user-specific K-factors
- Automatic sensor zero offset adjustment (via digital interface or local push button)
- NIST traceable certificate included
- Self-diagnostic tests
Application
- Gas flow measurement
- Gas glow control
- Operating pumps and valves
- Process equipment
- Vacuum processes
- Glass and metal coating
- Film deposition
Dwyer GFM2 Series Specification
Service | Clean gases compatible with wetted parts. |
Wetted Materials | GFM2-X-X-A: Anodized aluminum, brass, 316 SS fluoroelastomer O-rings; |
GFM2-X-X-S: 316 SS, and fluoroelastomer O-rings; Buna-N, EPR and PTFE O-rings optional. | |
Accuracy | ±1% FS. |
Repeatability | ±0.25% FS. |
Response Time | 2 seconds to within ±2% of actual flow. |
Output Signal | Linear 0 to 5 VDC (3000 Ω min. load impedance) and 4 to 20 mA (500 Ω max. loop resistance). |
Relay Rating | 1 A @ 24 VDC. |
Max. Particulate Size | 5 microns. |
Temperature Limits | 32 to 122°F (0 to 50°C). |
Power Supply | 11 to 26 VDC. |
Process Connections | 1/8" compression fitting for flow rates ≤10 L/min; 1/4" for ≤50 L/min; 3/8" for ≤100 L/min. |
Display | 2 x 16 character LCD (optional). |
Pressure Limits | 500 psig (34.5 bar). |
Leak Integrity | 1 x 10-9 smL/sec of helium. |
Weight | 1.05 lb (0.48 kg). |
Download Data Sheet: Series GFM2 Gas Mass Flow Meter and Controllers
Ref: dwyer-inst.com