| Dwyer TFP-GV Series Gas Turbine Flow Meter |
Dwyer TFP-GV Series Features
- Corrosion resistant PPS body
- High repeatability with patented microturbine design
- Design accurately measures flow rates with no zero drift and no required maintenance
Application
- Industrial flow monitoring
- Commercial systems
- Laboratory equipment
Dwyer TFP-GV Series Specification
| Service | Clean dry gases compatible with wetted materials. |
| Wetted Materials | PPS, acetal, sapphire, glass, epoxy, and fluoroelastomer. |
| Accuracy | ±3% of FS. |
| Linearity | ±3% of FS. |
| Repeatability | ±0.5% of FS. |
| Temperature Limits | 41 to 131°F (5 to 55°C); Storage: 32 to 158°F (0 to 70°C); Sensitivity: ±0.2% of FS per °C. |
| Pressure Limits | 40 psig (2.8 bar). |
| Process Connection | Compression fitting |
| Power Requirements | 11.5 to 15 VDC. |
| Power Consumption | 35 mA @ 12 VDC. |
| Output Signal | 0 to 5 VDC; minimum 2.5 kΩ load. |
| Electrical Connections | Four-pin power and signal connector. A power adapter or mating cable required for operation. |
| Enclosure Rating | IP10 (NEMA 1). |
| Weight | 0.16 lb (75 g). |
| Agency Approvals | CE. |
Download Data Sheet: Series TFP-GV Gas Turbine Flow Meter
Ref: dwyer-inst.com